COOK ENDOSCOPY ACROBAT¿ CALIBRATED TIP WIRE GUIDE; OCY, ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY
|
Back to Search Results |
|
| Catalog Number |
ACRO-35-260 |
| Medical Device Problem Code |
Peeled/Delaminated (1454)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); Foreign Body In Patient (2687)
|
| Date of Event |
05/18/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: cook fusion pre-loaded with acrobat wire guide (fs-omni-acro-35-260).Investigation evaluation: our evaluation of the product said to be involved confirmed the report.There is wire guide coating damage near the distal end.The coil spring is still attached to the distal end of the wire guide.The silver print markers appear to be stained yellow.There are a few kinks on the distal end of the device, particularly at approximately 6 cm.Approximately 6.7 cm to 15.9 cm from the distal end, the wire guide covering has folded over itself at least once.Approximately 8 cm from the distal end, 1 cm of wire guide coating is hanging off of the wire guide.Approximately 8.9 cm from the distal end, 2 mm of wire guide coating is hanging off of the wire guide.Approximately 15.9 cm to 26.8 cm from the distal end is a section of bare core wire.The wire guide coating feels rough between 154 cm and 171 cm from the distal end.Due to the condition of the returned device it cannot be determined if any sections of the coating are missing.A product-specific discrepancy that could have caused or contributed to this observation was not observed during our laboratory analysis.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.The instructions for use instructs the user to do the following: "prior to removing wire guide from holder, flush with 30 cc of sterile water." failure to flush the wire guide can result in damage to the wire guide.The instructions for use instructs the user to do the following: "flush endoscope accessory channel and/or lumen of device with sterile water, then insert wire guide floppy end first.Note: for best results, wire guide should be kept wet, if applicable." failure to flush the endoscope channel can result in damage to the wire guide.The instructions for use precautions the user that this product is not compatible with metal tip devices."use of this wire guide with metal tip ercp devices may result in damage to the external coating and/or tip of the wire guide." the reported observation can occur if the wire guide was used with an incompatible accessory device.If additional pressure is applied to the wire guide and/or accessory device(s) while moving the wire guide inside the accessory device(s), this could contribute to wire guide coating damage.Prior to distribution, all acrobat calibrated tip wire guides are subjected to a visual inspection to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
| |
|
Event or Problem Description
|
|
During an endoscopic retrograde cholangiopancreatography (ercp), the physician used a cook acrobat calibrated tip wire guide.The coating detached from the acrobat wire guide after cannulation which made exchange of devices impossible.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: added foreign, company representative, user facility, health professional.Investigation evaluation: our evaluation of the product said to be involved confirmed the report.There is wire guide coating damage near the distal end.The coil spring is still attached to the distal end of the wire guide.The silver print markers appear to be stained yellow.There are a few kinks on the distal end of the device, particularly at approximately 6 cm.Approximately 6.7 cm to 15.9 cm from the distal end, the wire guide covering has folded over itself at least once.Approximately 8 cm from the distal end, 1 cm of wire guide coating is hanging off of the wire guide.Approximately 8.9 cm from the distal end, 2 mm of wire guide coating is hanging off of the wire guide.Approximately 15.9 cm to 26.8 cm from the distal end is a section of bare core wire.The wire guide coating feels rough between 154 cm and 171 cm from the distal end.Due to the condition of the returned device it cannot be determined if any sections of the coating are missing.A product-specific discrepancy that could have caused or contributed to this observation was not observed during our laboratory analysis.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.The instructions for use instructs the user to do the following: "prior to removing wire guide from holder, flush with 30 cc of sterile water." failure to flush the wire guide can result in damage to the wire guide.The instructions for use instructs the user to do the following: "flush endoscope accessory channel and/or lumen of device with sterile water, then insert wire guide floppy end first.Note: for best results, wire guide should be kept wet, if applicable." failure to flush the endoscope channel can result in damage to the wire guide.The instructions for use precautions the user that this product is not compatible with metal tip devices."use of this wire guide with metal tip ercp devices may result in damage to the external coating and/or tip of the wire guide." the reported observation can occur if the wire guide was used with an incompatible accessory device.If additional or excess pressure is applied to the wire guide and/or accessory device(s) while moving the wire guide inside the accessory device(s), this could contribute to wire guide coating damage.Prior to distribution, all acrobat calibrated tip wire guides are subjected to a visual inspection to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
| |
|