• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDICAL INTELLIGENCE MEDIZINTECHNIK GMBH IGUIDE SYSTEM; ACCELERATOR, LINEAR, MEDICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDICAL INTELLIGENCE MEDIZINTECHNIK GMBH IGUIDE SYSTEM; ACCELERATOR, LINEAR, MEDICAL Back to Search Results
Model Number P10603-109 AND P10603-110
Medical Device Problem Code Computer Software Problem (1112)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer discovered during an in-house investigation that if the pec data dialog after a verification scan is cancelled, a re-send of pec data will lead to incorrect positioning because the 2nd pec values supersede the 1st pec values instead of being added.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IGUIDE SYSTEM
Common Device Name
ACCELERATOR, LINEAR, MEDICAL
Manufacturer (Section D)
MEDICAL INTELLIGENCE MEDIZINTECHNIK GMBH
robert-bosch-strasse 8
schwabmuenchen, 86830
GM  86830
MDR Report Key6633208
Report Number9710332-2017-00002
Device Sequence Number13455919
Product Code IYE
UDI-Device Identifier04056719001742
UDI-Public04056719001742
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K143485
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Service Personnel
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 06/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberP10603-109 AND P10603-110
Device Catalogue NumberP10603-109 AND P10603-110
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/19/2017
Initial Report FDA Received Date06/12/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-