| Model Number |
48811357 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Code Available (3191)
|
| Date of Event |
05/15/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that; surgeon was doing a 3-6 acdf with a 57mm plate.He initially put all 8 screws in 2/3 of the way, not using the drill guide.He tightened down the screws at the top level and bottom level, and when he went to tighten down the screws at the middle levels, he noticed one of the wings had snapped out of the plate.He backed out all the screws, damaging some of the other locking wings.A new 57mm plates was installed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Method: visual inspection; device history review; complaint history review; risk assessment; results: the returned plate was confirmed to have deformed locking mechanisms.Manufacturing files were reviewed and no anomalies were found.Conclusion: the root cause is not using the drill guides as recommended.
|
| |
|
Event or Problem Description
|
|
It was reported that; surgeon was doing a 3-6 acdf with a 57mm plate.He initially put all 8 screws in 2/3 of the way, not using the drill guide.He tightened down the screws at the top level and bottom level, and when he went to tighten down the screws at the middle levels, he noticed one of the wings had snapped out of the plate.He backed out all the screws, damaging some of the other locking wings.A new 57mm plates was installed successfully.
|
| |
|