| Model Number |
3 |
| Medical Device Problem Code |
Migration or Expulsion of Device (1395)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/14/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The event product was returned to applied medical for evaluation.Upon inspection, engineering confirmed the complainant's experience of a hole in the pouch.The root cause of the event is likely due to shipping and handling.The probability and criticality of harm resulting from this failure mode have been evaluated and were found to be at an acceptable level.Applied medical will continue to monitor its vigilance system for trends and take appropriate actions, as necessary to ensure the performance and safety of its products.
|
| |
|
Event or Problem Description
|
|
Procedure performed: na.Incident description: "we finished our receiving inspection for invoice #93902737.Some of the products did not pass the receiving inspection because of particulate matter and so on.Please find the attached rejected report for the detailed information.We would like applied medical to issue credit memo for the rejected items, but we don't need credit memo only for g6000 (lot#1274065)." type of intervention: na.Patient status: na.
|
| |
|