| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/22/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The account generated a falsely depressed architect glucose of 10 mg/dl on a sample that repeated 143 mg/dl on another analyzer.The sample was processed again on the original analyzer with architect glucose of 145 mg/dl.No specific patient information or impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Review of complaint activity determined there were no other complaints for likely cause lot 49673uq07.Tracking and trending report review for the architect glucose assay determined that there are no related adverse or non-statistical trends.Manufacturing release documents were reviewed and all specifications were met prior to release.Labeling was reviewed and sufficiently addresses the customer's issue.No systemic issue or deficiency of the architect glucose assay was identified.
|
| |
|