Information was received from a healthcare provider (hcp) regarding a patient who was receiving 300 mcg/ml of clonidine at 108.19 mcg/day and 20 mg/ml of morphine at 7.213 mg/day via an implantable pump for non-malignant pain.On (b)(6) 2017, it was reported that a pump motor stall occurred when the patient had an mri on (b)(6) 2017 and the stall had not recovered.The hcp reported that the patient had an mri on (b)(6) 2017 at 11:33 and, during a refill the following day, it was noted that the motor stall had occurred.The hcp confirmed that the motor stall occurred on (b)(6) 2017 at 11:33, the same time the patient was having the mri.The pump alarm was heard after the pump was interrogated.According to the hcp, no recovery was noted both 10 and 15 minutes after the initial interrogation.No symptoms were reported and no further complications were reported.
|