An end-of-service warning message was verified in the lab for the explanted generator and found to be associated with the output being disabled by the pulse generator.Various electrical loads were attached to the pulse generator and results of diagnostic tests demonstrate that accurate resistance measurements were obtained in all instances.A comprehensive automated electrical evaluation showed that the pulse generator performed according to functional specifications.A large portion of the lead assembly including the electrodes was not returned for analysis; therefore a complete evaluation could not be performed on the entire lead product.What appeared to be white deposits were observed in various locations.The condition of the returned lead portions is consistent with conditions that typically exist following an explant procedure.No obvious anomalies were noted.The set screw marks found on the lead connector pins provide evidence that, at one point in time, a good mechanical and electrical connection was present.Continuity checks of the returned lead portions were performed, during the visual analysis, and no discontinuities were identified.Based on the findings in the lab, there is no evidence to suggest discontinuities in the returned portions of the device.
|