MAUDE Adverse Event Report: RESPIRONICS, INC. TRILOGY 202; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. TRILOGY 202; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 1040007
Medical Device Problem Code Incorrect Or Inadequate Test Results (2456)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/18/2017
Type of Reportable Event Malfunction
Event or Problem Description
A ventilator was returned to the manufacturer for routine preventive maintenance.There was no allegation of device malfunction.The device was not in patient use.During the evaluation of the device at the manufacturer's service center, the device failed steps during testing.The device's oxygen blending module board, oxygen blender module gasket, oxygen blender module manifold were replaced to address the issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY 202
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
MDR Report Key6639653
Report Number2518422-2017-01367
Device Sequence Number19103251
Product Code CBK
UDI-Device Identifier00606959026247
UDI-Public00606959026247
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K093905
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Service and Testing Personnel
Type of Report Initial
Report Date (Section B) 05/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1040007
Device Catalogue Number1040007
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/18/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/18/2017
Initial Report FDA Received Date06/14/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-