Catalog Number DBF-2.4SN-230SP-40-S
Medical Device Problem Code Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/23/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation evaluation: our laboratory evaluation of the product said to be involved confirmed the report.During the evaluation of the returned device, the device was functionally tested.When the handle of the device was manipulated the forceps cups will open, but they will not close.Several attempts were made to close the forceps cups, but they were unsuccessful.The device was then placed down an olympus gif-q20 (2.8 mm channel) endoscope.The endoscope was put in a torturous path.When the handle was manipulated the device would open, but the cups would not close.The device was sent back to the supplier for evaluation.The supplier provided the following evaluation: one device from the reported event was returned in a zip type bag with proof of decontamination.The returned device from the reported event was tested for "would not close".During functional testing, it was confirmed that the device would open but would not close when manipulated using the handle.Upon disassembly of the device tip, it was noted that the solder joint was broken.The reported defect was confirmed for the device.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: the customer experienced issue of "would not close" was confirmed; the device opens but does not close.The device has a butt joint that has broken at the solder connection.The root cause is a butt joint with a broken control wire / link wire solder joint.The supplier is currently working on improvements to create a more robust soldering process to prevent solder joints from breaking.The instructions for use state the following in regards to product inspection: "beginning at the handle and moving toward the cups, uncoil the forceps making sure not to stretch the cable.Open and close the cups to verify smooth handle operation and appropriate cup action.Become familiar with the amount of handle movement required to operate the cups.If any irregularities are noted, do not use.Note: exercising the handle while the forceps is coiled may result in damage to the performance characteristics of the forceps." prior to distribution, all captura serrated forceps with spike are subjected to a visual inspection and functional test to ensure proper workability.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event or Problem Description
During a colonoscopy, the physician used a cook captura serrated forceps with spike."during a colonoscopy procedure, the physician did one biopsy and was attempting to take the second.When the forceps were opened something snapped and the forceps would not close.[the device] was removed from the endoscope in an opened position and another device was used to complete the procedure.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTURA SERRATED FORCEPS WITH SPIKE
Common Device Name
FCL, FORCEPS, BIOPSY, NON-ELECTRIC
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
MDR Report Key6639671
Report Number1037905-2017-00385
Device Sequence Number19103295
Product Code FCL
UDI-Device Identifier00827002561335
UDI-Public(01)00827002561335(17)200322(10)W3844072
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 06/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberDBF-2.4SN-230SP-40-S
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date05/23/2017
Device Age2 MO
Event Location Outpatient Treatment Facility
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2017
Initial Report FDA Received Date06/14/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OLYMPUS TF 190 ENDOSCOPE
-
-