| Model Number |
M00560160 |
| Medical Device Problem Code |
Failure to Deliver Energy (1211)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/18/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an injection gold probe was used in the upper gi during a hemostasis procedure performed on (b)(6) 2017.According to the complainant, during the procedure, the injection gold probe failed to cauterized.The procedure was completed with another injection gold probe.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results: visual evaluation of the returned device found no visible failures.An electrical load test was performed and the device tested to be within specification.Based on all gathered information and investigation results, there is no evidence of either the alleged issue or any defect which could have contributed to the event.A review of the device history record (dhr) was performed and no deviations were found.A search of the complaint database revealed that no other complaints exist for the specified lot.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an injection gold probe was used in the upper gi during a hemostasis procedure performed on (b)(6) 2017.According to the complainant, during the procedure, the injection gold probe failed to cauterized.The procedure was completed with another injection gold probe.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
|
| |
|