|
Device evaluation summary: the returned specimen consists of one-1 each mjncca-x, int, ptfe 260-035; returned still loaded within a trailblazer support catheter.The device presents bend damage at 2.02 to 3.31cm from the distal tip, and pinch damage to the ptfe sleeve 76.8 to 79.6cm from the distal tip.The pinch damage presents witness impressions consistent with a torque device or similar constraint device and results in an out of round condition.The proximal 0.64mm of the ptfe sleeve present tear damage.No other damage or inconsistencies are noted to the specimen.All joints appear to be correct and intact by visual examination and by non-destructive testing.Except where noted, the specimen device appears visually and dimensionally correct.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|