| Model Number |
840 |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
05/26/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that an 840 ventilator had a faulty display.The ventilator was not on a patient at the time of the event.The service engineer replaced the backlight inverter board, the gui (graphical user interface) board and downloaded software to correct the issue.The unit passed all testing and operates within the manufacturing specifications.
|
| |
|
Search Alerts/Recalls
|