| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has been returned to animas.Evaluation has not yet been completed.When evaluation is complete a supplemental report will be filed.No conclusion can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2017, the reporter contacted animas, alleging a power (battery life) issue.There was no allegation of an adverse event associated with this complaint.This complaint is being reported because the user may be unaware that the pump has lost power, leading to under delivery.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device has been returned and evaluated by product analysis on 06/28/2017 with the following findings:.No defect was found.Black box shows no evidence of short battery life, returned battery cap is able to fit to maintain electrical connection ¿ez-prime¿ steps were performed correctly, all currents reading are within specification , unable to duplicate ¿short battery life¿ complaint.Pump¿s cover was removed, no intermittent condition was found to the printed circuit board.
|
| |
|
Search Alerts/Recalls
|