Analysis: the sample was not returned to the user facility; therefore, device evaluation is unable to be performed.A lot history review revealed there are no similar complaint associated with this lot.A review of the device history record (dhr) shows the lot was manufactured to specification.Conclusion: the actual sample was not received for evaluation.The dhr found nothing to indicate a manufacturing related cause for this event.The investigator original assessed the patient's target lesion was patent but after the independent evaluation was provided the site changed their assessment to reoccluded.Based on the instructions for use (ifu), the occurrence of a reocclusion is an inherent risk of any pta procedure, and has been reported in clinical trials of drug coated balloons.If additional information becomes known to the manufacturer, a supplemental report will be provided will all relevant information.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
|
It was reported through a clinical registry that during the index procedure, a lutonix percutaneous transluminal drug coated balloon (dcb) dilatation catheter was used to treat the target lesion located in the right superficial femoral artery (sfa).Approximately 11 months after the index procedure, the patient¿s right sfa vessel was reportedly reoccluded.No additional intervention has been performed at this time.The registry's independent evaluator's have determined the patient developed reocclusion in approximately 11 months after index procedure.The investigator original assessed the patient's target lesion was patent but after the independent evaluation was provided, the site changed their assessment to reoccluded.The sample was discarded by the user facility and not available for further evaluation.No adverse patient effects were reported.During further follow-up, it was discovered the patient underwent a reintervention.
|