• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC COUGAR GUIDE WIRE; WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC COUGAR GUIDE WIRE; WIRE, GUIDE, CATHETER Back to Search Results
Catalog Number CGRXT190HS
Medical Device Problem Code Material Distortion (2977)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/02/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product analysis: the guide wire exhibited two stretched sections.The stretched coils were noted approximately 3cm and 1.5cm from the distal tip ball.The distal tip ball was intact.No other damage was noted to the guide wire.Physical measurements show the device meets specification.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The physician intended to use a cougar guide wire during a procedure.No damage was noted to the device packaging and no issues noted when removing the wire from the hoop.The wire was inspected and prepped per ifu with no issues noted.Damage was noted to the wire on removal from the patient.Analysis of the returned wire identified stretched coils.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COUGAR GUIDE WIRE
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
MEDTRONIC, INC
37a cherry hill dr
danvers MA 01923
MDR Report Key6642795
Report Number1220452-2017-00056
Device Sequence Number16388480
Product Code DQX
UDI-Device Identifier20643169453320
UDI-Public20643169453320
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K032899
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/16/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/20/2019
Device Catalogue NumberCGRXT190HS
Device Lot NumberG16A06130
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/21/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/16/2017
Initial Report FDA Received Date06/15/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-