| Catalog Number |
CGRXT190HS |
| Medical Device Problem Code |
Material Distortion (2977)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/02/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product analysis: the guide wire exhibited two stretched sections.The stretched coils were noted approximately 3cm and 1.5cm from the distal tip ball.The distal tip ball was intact.No other damage was noted to the guide wire.Physical measurements show the device meets specification.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The physician intended to use a cougar guide wire during a procedure.No damage was noted to the device packaging and no issues noted when removing the wire from the hoop.The wire was inspected and prepped per ifu with no issues noted.Damage was noted to the wire on removal from the patient.Analysis of the returned wire identified stretched coils.
|
| |
|