Model Number INFKIT2
Device Problem Defective Alarm (1014)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/04/2017
Event Type  malfunction  
Manufacturer Narrative
The pump was returned to mmdg for evaluation.During visual inspection of the pump, corrosion was found due to fluid ingress.Due to the fluid ingress the pump would not turn on and could not be tested for the alleged complaint.The pump was repaired and returned to the customer.
 
Event Description
The initial reporter stated that the pump did not alarm when the flow was obstructed.They also stated that this occurred during testing, and that there was no patient involvement.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERALITE INFINITY ENTERAL FEEDING PUMP
Type of Device
ENTERAL INFUSION PUMP
Manufacturer (Section D)
MOOG MEDICAL DEVICES GROUP
4314 zevex park lane
salt lake city UT 84123
Manufacturer (Section G)
MOOG MEDICAL DEVICES GROUP
4314 zevex park lane
salt lake city UT 84123
Manufacturer Contact
kristin hardesty
4314 zevex park lane
salt lake city, UT 84123
8012641001
MDR Report Key6643491
MDR Text Key77646047
Report Number1722139-2017-00089
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031199
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 05/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberINFKIT2
Device Catalogue NumberINFKIT2
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/19/2017
Initial Date FDA Received06/15/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/04/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
NOT PROVIDED
-
-