APPLIED MEDICAL RESOURCES CTR71, KII 12X150 Z-THR OPT SEP SYS 6/BX; GCJ Back to Search Results
Model Number 4
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/23/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The event device is anticipated to return.A follow-up report will be provided upon completion of the investigation.
 
Event or Problem Description
Procedure performed: turp with da vinci robotic surgery.Event description: "in surgery, this device was used with endoscope but it was broken in patient body suddenly." photograph provided.Type of intervention: no patient injury or illness occurred associated with the event patient status: "fine".
 
Additional Manufacturer Narrative
The event product was returned to applied medical for evaluation.Upon inspection, engineering was able to confirm the complainant's experience of a fractured cannula.The wall thickness of the cannula was measured and was found to be uneven.The likely root cause of the fracture is the uneven cannula wall thickness, which occurred during the injection molding process of the cannula.The probability and criticality of harm resulting from this failure mode have been evaluated and were found to be at an acceptable level.Applied medical continuously seeks to improve the form, function and ease of use of its products.As part of this process, applied medical is currently researching possible process enhancements intended to further minimize the potential for this type of incident to occur.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CTR71, KII 12X150 Z-THR OPT SEP SYS 6/BX
Common Device Name
GCJ
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
MDR Report Key6644186
Report Number2027111-2017-01858
Device Sequence Number13656948
Product Code GCJ
UDI-Device Identifier00607915125042
UDI-Public(01)00607915125042(17)190815(30)01(10)1276446
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
4
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date08/15/2019
Device Model Number4
Device Catalogue Number4
Device Lot Number1276446
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/20/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/08/2017
Supplement Date Received by Manufacturer06/08/2017
Initial Report FDA Received Date06/15/2017
Supplement Report FDA Received Date07/13/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DAVINCI ROBOT; ENDOSCOPE
-
-