| Model Number |
48004004S |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/29/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
It was reported that a saber balloon catheter (bc) ruptured at five atmospheres (5 atm).Therefore it was replaced with a non cordis balloon catheter and the procedure was completed successfully.There was no reported patient injury.The target lesion was the left popliteal artery.The patient¿s vessel level of tortuousness and calcification was unknown.The rate of stenosis was unknown.The device inflated normally.There were no difficulties removing the product from the hoop, the protective balloon cover, the stylet or any of the sterile packaging components.There were no kinks or other damages noted prior to inserting the product the product into the patient.A non-cordis guidewire and a non cordis micro catheter were inserted and an attempt was made to inflate a saber percutaneous transluminal angioplasty balloon catheter.However it ruptured.The product was removed intact (in one piece) from the patient.No other information was provided.The device was not returned for analysis.A device history record (dhr) review of lot 17029240 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon burst - at/below rbp¿ could not be confirmed as the device was not returned for analysis.The exact cause of the burst reported could not be determined.Based on the limited information available for review, vessel characteristics (although unknown) may have contributed to the reported event.According to the instructions for use, which is not intended as a mitigation, ¿the rated burst pressure is based on the results of in vitro testing.At least 99.9% of the balloons (with a 95% confidence) will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over-pressurization.¿ neither the dhr nor the information available suggests a design or manufacturing related cause for the reported event; therefore, no corrective/preventive action will be taken.
|
| |
|
Event or Problem Description
|
|
It was reported that a saber balloon catheter ruptured at five atmosphere (5 atm).Therefore it was replaced with a non cordis balloon catheter and the procedure was completed successfully.There was no reported patient injury.A guidewire (chevalier, fmd) and a non cordis micro catheter were inserted and an attempt was made to inflate a saber percutaneous transluminal angioplasty balloon catheter.However it ruptured.The target lesion was the left popliteal artery.The patient¿s vessel level of tortuousness and calcification was unknown.The rate of stenosis was unknown.The device inflated normally.There were no difficulties removing the product from the hoop, the protective balloon cover, the stylet or any of the sterile packaging components.There were no kinks or other damages noted prior to inserting the product the product into the patient.The product was removed intact (in one piece) from the patient.The product was clinically used and will not be returned for analysis.No other information was provided.
|
| |
|
Search Alerts/Recalls
|
|
|