• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL, L.P. TURON SHOULDER; HEAD, HUMERAL, OFFSET, 50-26

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL, L.P. TURON SHOULDER; HEAD, HUMERAL, OFFSET, 50-26 Back to Search Results
Model Number 520-50-126
Health Effect - Clinical Code Damage to Ligament(s) (1952)
Date of Event 05/16/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to a rotator cuff tear and needing to be converted from a primary shoulder to a reverse shoulder.There were no patient issues, the case went great.
 
Additional Manufacturer Narrative
The reason for this revision surgery was the rotator cuff tore.The previous surgery and the revision detailed in this investigation occurred 1 year and 6 months apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to the rotator cuff failure.There were no findings during this investigation that indicate that the reported device was the source or had a direct connection with the event.There are multiple factors that may contribute to the event that are outside the control of djo surgical are patient activities, excessive load or torque, poor bone quality, inadequate soft tissue support or trauma.No additional information was submitted with the complaint regarding pre-existing conditions of the patient or any activities that may have contributed to the event and hence a definitive root cause cannot be determined.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURON SHOULDER
Common Device Name
HEAD, HUMERAL, OFFSET, 50-26
Manufacturer (Section D)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin TX 78758
MDR Report Key6645113
Report Number1644408-2017-00474
Device Sequence Number13723703
Product Code KWS
UDI-Device Identifier00888912025072
UDI-Public(01)00888912025072
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K080402
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/29/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2019
Device Model Number520-50-126
Device Catalogue Number520-50-126
Device Lot Number938C1013
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 05/16/2017
Supplement Date Received by Manufacturer06/30/2017
Initial Report FDA Received Date06/15/2017
Supplement Report FDA Received Date07/29/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
-
-