| Catalog Number |
SPS-21120-35 |
| Health Effect - Clinical Code |
Vascular Dissection (3160)
|
| Date of Event |
05/25/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Visual inspection of the returned device noted buckling of the distal portion of the catheter which is consistent with the report that the device was advanced against resistance into an area of vasospasm.The device instructions for use includes the following: "caution: do not advance or torque the system against resistance.If the cause of resistance cannot be determined, carefully withdraw catheter.".
|
| |
|
Event or Problem Description
|
|
The physician was planning to use a surefire precision infusion system to perform a lower lobe kidney embolization through the renal artery.The physician tracked out to the desired location with a boston scientific renegade stc microcatheter and fathom 016 guide wire.The renegade microcatheter was removed and exchanged for the surefire precision infusion system which was tracked over the guidewire.When the surefire precision infusion system exited the guide catheter, arterial vasospasm was encountered.The physician attempted to advance the surefire precision infusion system through the area of vasospasm.This resulted in dissection of the vessel.Physician intervention was required to stop the bleeding and remove thrombosis from the artery.The patient was admitted to the hospital overnight for monitoring and released the following day.
|
| |
|