• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS EMERALD; SURGICAL ADJUNCTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS EMERALD; SURGICAL ADJUNCTS Back to Search Results
Model Number EMERALDC30
Device Problems Scratched Material (3020); Appropriate Term/Code Not Available (3191)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/08/2017
Event Type  malfunction  
Manufacturer Narrative
This is not an implantable device.(b)(6).(b)(4).Device evaluation: the cartridge was returned at the manufacturing site for evaluation.Visual inspection at 10x microscope magnification showed residue of viscoelastic solution on the cartridge, indicating that the unit was handled and prepare for surgical use.Dent/distortion was observed at the cartridge tip.The customer's reported complaint was verified.Manufacturing records review: the manufacturing records for the cartridge were reviewed.During the manufacturing process the operators check the neck, tube and tip areas for cracks.No cracking or stress marks are allowed.They also check the tip for any melting, roughness, dent, bent tip or smash condition.The product was manufactured and released according to specification.A search revealed that no additional investigation requests for this order number have been received.Labeling review: the directions for use (dfu) were reviewed.The directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to abbott medical optics has been submitted.
 
Event Description
It was reported that an intraocular lens (iol) was implanted in the eye of a (b)(6) male patient and the cartridge was noticed to have a marking and burr edge.No impact to the patient.Additional information revealed that the issue was located on the cartridge tip.No further information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMERALD
Type of Device
SURGICAL ADJUNCTS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
Manufacturer (Section G)
ABBOTT MEDICAL OPTICS INC.
road 402 north, km 4.2
anasco industrial park, pob 14
anasco PR 00610
Manufacturer Contact
pam mcclain
1700 east st. andrew place
santa ana, CA 92705
7142478243
MDR Report Key6645736
MDR Text Key77744592
Report Number2648035-2017-01074
Device Sequence Number1
Product Code KYB
UDI-Device Identifier05050474530027
UDI-Public(01)05050474530027(17)170727(10)CB36717
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
K961242
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 06/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date07/27/2017
Device Model NumberEMERALDC30
Device Catalogue NumberEMERALDC30
Device Lot NumberCB36717
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/01/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/22/2017
Initial Date FDA Received06/15/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/27/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age58 YR
-
-