• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL, L.P. FMP HIP; FMP HEMISPHERICAL SHELLS W/SCREW HOLES, 56MM, W/P2 COATING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL, L.P. FMP HIP; FMP HEMISPHERICAL SHELLS W/SCREW HOLES, 56MM, W/P2 COATING Back to Search Results
Model Number 932-40-256
Health Effect - Clinical Code Pain (1994)
Date of Event 05/18/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - the patient had a total hip arthoplasty (tha) performed in 2016 and since has been presenting symptoms of pain.The surgeon decided to replace the acetabular component and head.
 
Additional Manufacturer Narrative
The reason for this revision surgery was hip pain.The in-vivo length of patient service for the implant was 1.2 years.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was a significant adverse event to the patient.There was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.Initial or prolonged hospitalization was required.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the device history records (dhrs) revealed no discrepancies or issues with the manufacturing of this part.All critical dimensions, design criteria and specifications in effect at the time the part was manufactured were met.The product complaint report history was reviewed and no trends or on-going issues were deemed as present or in need of review.This event is deemed as non-product related.The root cause for this event was the patient had a hip pain.The root cause for the pain was not reported.The scope of this investigation is limited without having the explanted parts available to djo surgical for evaluation.Other conditions relating to this event could not be determined with confidence.Inventory containment is not required since there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMP HIP
Common Device Name
FMP HEMISPHERICAL SHELLS W/SCREW HOLES, 56MM, W/P2 COATING
Manufacturer (Section D)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin TX 78758 5445
MDR Report Key6648649
Report Number1644408-2017-00504
Device Sequence Number1306422
Product Code LPH
UDI-Device Identifier00888912100250
UDI-Public(01)00888912100250
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K974093
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2021
Device Model Number932-40-256
Device Catalogue Number430-98-056
Device Lot Number591G1106
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 05/18/2017
Supplement Date Received by Manufacturer06/29/2017
Initial Report FDA Received Date06/16/2017
Supplement Report FDA Received Date07/28/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
400-03-401,LOT 870B1059
Outcome Attributed to Adverse Event Hospitalization; Other; Required Intervention;
Patient Age60 YR
-
-