| Model Number |
ZCT225 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Visual Disturbances (2140); No Code Available (3191)
|
| Date of Event |
05/09/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).All pertinent information available to the manufacturer has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a zct225.20.5 diopter was explanted from a female patient's left eye due to varied refractive outcome.According to the physician, the patient experienced visual disturbance, anisometropia and pseudo phakia.The lens was replaced with the same model lens with a lower diopter (19.0).There was no incision enlargement, vitrectomy or sutures required.No patient injury reported.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation ¿ yes, returned to manufacturer on 07/03/2017.Device returned to manufacturer ¿ yes.Device evaluation: the intraocular lens (iol) was returned at the manufacturing site for evaluation.Visual inspection with the unaided eye shows the product is cut in pieces, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints were received for this production order number.Labeling review: the directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to abbott medical optics has been submitted.
|
| |
|
Search Alerts/Recalls
|