MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL DELTAPAQ - CERECYTE MICROCOIL; NEUROVASCULAR EMBOLIZATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL DELTAPAQ - CERECYTE MICROCOIL; NEUROVASCULAR EMBOLIZATION DEVICE Back to Search Results
Catalog Number CDF10051530
Medical Device Problem Codes Difficult To Position (1467); Stretched (1601)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/18/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).The device was returned for analysis; however, the analysis has not yet been completed.Additional information will be submitted within 30 days of receipt.
 
Event or Problem Description
As reported by a healthcare professional, during the procedure, the deltapaq coil (cdf10051530/ c39413) could not be inserted into the aneurysm due to positioning difficulty and stretched.They withdrew the coil without difficulty, and used a new coil to complete the procedure.There had been no resistance between the deltapaq and the microcatheter, and a continuous flush had been maintained through the microcatheter.Aneurysm characteristics and size were unavailable.There was no report of patient injury or procedure delay.It was reported that the device would be returned for analysis.
 
Additional Manufacturer Narrative
As reported by a healthcare professional, during the procedure, the deltapaq coil (cdf10051530/ c39413) could not be inserted into the aneurysm due to positioning difficulty and stretched.They withdrew the coil without difficulty, and used a new coil to complete the procedure.There had been no resistance between the deltapaq and the microcatheter, and a continuous flush had been maintained through the microcatheter.Aneurysm characteristics and size were unavailable.There was no report of patient injury or procedure delay.It was reported that the device would be returned for analysis.The device was returned in the packaging hoop within the opened inner pouch.The device could not be removed from the introducer hoop; therefore, the hoop was dissected and the device positioning unit (dpu) core wire along with the hub of the device and the embolic coil were separated from its introducer sheath.There were no damages observed on the dpu core wire.No damages were found on the embolic coil and the ball tip is intact at the distal end of the coil.The coil did not appear stretched.The complaint that the coil stretched and was difficult to position was not confirmed.There were no damages or stretches found on the dpu or embolic coil that could cause the issues.As the introducer sheath was removed from the rest of the unit, the user would not be able to resheath the device; however, this would not cause the positioning difficulty.The cause of positioning difficulty as reported is unknown; however, positioning can be influenced by the compatibility of the selected coil with the aneurysm¿s size and neck width.There is no current safety signal identified related to the reported event based on review of complaint history for the device.Since there was no evidence to suggest the event was related to a manufacturing issue, no corrective actions will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTAPAQ - CERECYTE MICROCOIL
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel
SZ 
MDR Report Key6650488
Report Number2954740-2017-00173
Device Sequence Number926358
Product Code HCG
UDI-Device Identifier00878528007302
UDI-Public(01)00878528007302(17)201031(10)C39413
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K080437
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 05/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2020
Device Catalogue NumberCDF10051530
Device Lot NumberC39413
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2017
Initial Date Received by Manufacturer 05/27/2017
Supplement Date Received by Manufacturer08/24/2017
Initial Report FDA Received Date06/19/2017
Supplement Report FDA Received Date08/24/2017
Patient Sequence Number1
-
-