• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM Back to Search Results
Model Number MT22495-PNK
Device Problem Failure to Power Up (1476)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/04/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Dexcom was made aware on (b)(6) 2017 that on (b)(6) 2017, that the receiver would not charge.No additional event or patient information is available.The complaint device was returned for evaluation.The device was visually inspected and no defect was found.Functional testing was performed and no failures were observed related to the customer complaint.The receiver log was reviewed and a screen error alert and a hardware error were observed.The customer complaint was confirmed during log review.The root cause was determined to be a hardware error and screen error alert.Based on the investigation results this issue has been upgraded from non-reportable to reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key6651995
MDR Text Key77939674
Report Number3004753838-2017-38747
Device Sequence Number1
Product Code MDS
UDI-Device Identifier10386270000160
UDI-Public(01)10386270000160(241)STK-DR-PNK(10)5219489
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 05/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMT22495-PNK
Device Catalogue NumberSTK-DR-PNK
Device Lot Number5219489
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/18/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/25/2017
Initial Date FDA Received06/19/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/26/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age9 YR
Patient Weight37
-
-