| Model Number |
48285545 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Injury (2348); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/03/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that; screw lost the polyaxiality.In the intraoperative moment, the rigidity of the screw was identified.
|
| |
|
Additional Manufacturer Narrative
|
|
Method: visual inspection, functional inspection, device history review, complaint history review, risk assessment.The screw was returned with some markings and deformation however it did not affect screw functionality.Per functional test, the screw functions as intended.Manufacturing history was reviewed and no issues were identified.The device functioned as expected, the reported event could not be duplicated.
|
| |
|
Event or Problem Description
|
|
It was reported that; screw lost the polyaxiality.In the intraoperative moment, the rigidity of the screw was identified.
|
| |
|