| Model Number |
100208H2HS-V-A2 |
| Medical Device Problem Codes |
Detachment Of Device Component (1104); Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/22/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The lot number was provided.A review of the approved device history records indicated the lot met all release criteria.The device was returned for evaluation.The pusher heater coil was observed to be compressed and deformed and the implant coil was detached; however, the root cause could not be determined.
|
| |
|
Event or Problem Description
|
|
It was reported that during a coil embolization treatment of a para-opthalmic aneurysm, the coil unexpectedly detached in the hub of the microcatheter.Both segments of the coil were removed in their entirety.There was no reported intervention or patient injury.The patient's current status is reported to be fine.
|
| |
|