| Catalog Number |
42546 |
| Medical Device Problem Codes |
Inaccurate Flow Rate (1249); Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/19/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product testing is in progress.A review of the manufacturing records showed no anomalies.All valves are 100% tested at the time of manufacture.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported to medtronic neurosurgery that when the nurse did the routine device test of the valve prior to implant, they found that it was not working.According to the report, the valve was replaced and there was no injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information received reported the valve would not drain normally.The returned valve was patent and met the requirements for reflux and leak testing.Therefore the conditions of the complaint could not be duplicated by laboratory personnel.However, the valve did not meet the requirements for pressure-flow and preimplantation testing.There was proteinaceous debris observed within the interior and exterior of the valve.The instructions for use that accompany the device caution that ¿shunt obstruction may occur in any of the components of the shunt system.The system may become occluded internally due to tissue fragments, blood clots, tumor cell aggregates, bacterial colonization, or other debris.¿ a review of the manufacturing records showed no anomalies.All valves are 100% tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|