| Model Number |
ZCB00 |
| Medical Device Problem Codes |
Device Damaged Prior to Use (2284); Scratched Material (3020)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
04/03/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If implanted, give date: not applicable no patient contact reported.If explanted, give date: not applicable no patient contact reported.All pertinent information available to abbott medical optics has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that that a scratch was noticed on an intraocular lens (iol) upon removing the lens from the package.No patient contact or injury was reported.Another amo lens was used to complete the procedure.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation? yes.Returned to manufacturer on: 07/14/2017.Device returned to manufacturer? yes.Device evaluation: the intraocular lens (iol) was returned at the manufacturing site for evaluation.Visual inspection at 10x microscope magnification showed loose particles on the lens surface.Also, some white residue was observed attached to the lens posterior surface.However, no scratches, defects or damages related to the manufacturing process were observed on the lens.The customer's reported complaint was not verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.During manufacturing the operators clean and visually inspect 100% of the lenses at 10x microscope magnification.The product was manufactured and released according to specification.A search revealed that no additional complaints for this order number have been received.Labeling review: the directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.The dfu states to examine the lens thoroughly to ensure particles have not become attached to it, and examine the lens optical surfaces for other defects.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to abbott medical optics has been submitted.
|
| |
|
Search Alerts/Recalls
|