ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Medical Device Problem Code Structural Problem (2506)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The pump has been returned to animas.Evaluation has not yet been completed.When evaluation is complete a supplemental report will be filed.No conclusion can be made at this time.
 
Event or Problem Description
On (b)(6) 2017, the reporter contacted animas, alleging a casing/condition (damaged cap and case) issue.The reporter alleged that the battery cap was unable to be removed.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the alleged malfunction has the ability to result in a delay in treatment or long term cessation in delivery if the damage impacts the power circuit or the cartridge compartment.
 
Additional Manufacturer Narrative
Device evaluation: the device has been returned and evaluated by product analysis on 06/29/2017 with the following findings:.The battery compartment was found to be free of defects.The damaged battery compartment case complaint could not be confirmed.However, the battery cap was observed to have damaged threads, and was difficult to remove from the complainant pump.Additionally, the height of the battery cap contact was out of specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Common Device Name
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
MDR Report Key6655979
Report Number2531779-2017-12932
Device Sequence Number19098417
Product Code OYC
UDI-Device Identifier10840406100310
UDI-Public0110840406100310
Combination Product (Y/N)N
Initial Reporter StateSD
Initial Reporter CountryUS
PMA/510(K) Number
P130007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Patient
Type of Report Initial,Followup
Report Date (Section B) 05/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2017
Was the Report Sent to FDA? Yes
Initial Date Received by Manufacturer 05/26/2017
Supplement Date Received by Manufacturer05/26/2017
Initial Report FDA Received Date06/20/2017
Supplement Report FDA Received Date07/26/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age8 YR
-
-