MAUDE Adverse Event Report: ARTHREX, INC. LARGE PEC BUTTON KIT; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. LARGE PEC BUTTON KIT; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number AR-2269
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 03/16/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Patient demographics (age at time of event, gender, weight) were requested but not provided.No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.Device disposition is unknown but has been requested.If available for evaluation return will be requested.The complainant's event could not be verified.The cause of the event could not be determined from the information available and without device evaluation.Device history record review revealed nothing relevant to this event.The most likely cause(s) of this type of event is prying/leveraging the driver while the implant is still loaded.This is the first complaint of this type for this part/lot combination.The potential cause(s) of this event will be communicated to the event reporter.If the additional relevant information is received, a follow-up report will be submitted.
 
Event or Problem Description
It was reported via medwatch (b)(4) that during a pectoralis major tendon repair the large pec button kit was used.After insertion of the pec major large eyelet into the humerus using a pec repair implant delivery system, it was noticed that one of the stylets used in the process was missing a threaded portion.The missing portion measured approximately 2 x 1 mm.The wound was explored and the fragment was not found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LARGE PEC BUTTON KIT
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
MDR Report Key6656266
Report Number1220246-2017-00218
Device Sequence Number19202016
Product Code MBI
UDI-Device Identifier00888867026582
UDI-Public00888867026582
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K123341
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Risk Manager
Type of Report Initial
Report Date (Section B) 06/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/01/2021
Device Catalogue NumberAR-2269
Device Lot Number10059387
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/30/2017
Initial Report FDA Received Date06/21/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age50 YR
Patient Weight118
-
-