| Catalog Number |
AR-2269 |
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
03/16/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Patient demographics (age at time of event, gender, weight) were requested but not provided.No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.Device disposition is unknown but has been requested.If available for evaluation return will be requested.The complainant's event could not be verified.The cause of the event could not be determined from the information available and without device evaluation.Device history record review revealed nothing relevant to this event.The most likely cause(s) of this type of event is prying/leveraging the driver while the implant is still loaded.This is the first complaint of this type for this part/lot combination.The potential cause(s) of this event will be communicated to the event reporter.If the additional relevant information is received, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported via medwatch (b)(4) that during a pectoralis major tendon repair the large pec button kit was used.After insertion of the pec major large eyelet into the humerus using a pec repair implant delivery system, it was noticed that one of the stylets used in the process was missing a threaded portion.The missing portion measured approximately 2 x 1 mm.The wound was explored and the fragment was not found.
|
| |
|