Model Number 1039363
Medical Device Problem Code Fire (1245)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/22/2017
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging an everflo oxygen concentrator caught on fire.There was no report of patient harm or injury.The device was returned to the manufacturer's service center for evaluation.The customer's complaint was confirmed.The device had evidence of thermal damage to the outlet port.The thermal damage was caused by an external source.The device was tested and was found to operate and alarm to design specifications.Product labeling states, "oxygen vigorously accelerates combustion and should be kept away from heat or open flame.Not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.Do not smoke, allow others to smoke or have open flames near the concentrator when it is in use.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERFLO OXYGEN CONCENTRATOR
Common Device Name
GENERATOR, OXYGEN, PORTABLE
Manufacturer (Section D)
RESPIRONICS
1001 murry ridge lane
murrysville PA 15668
MDR Report Key6657410
Report Number1040777-2017-00019
Device Sequence Number2064473
Product Code CAW
UDI-Device Identifier00606959032675
UDI-Public00606959032675
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K061261
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date (Section B) 05/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1039363
Device Catalogue Number1039363
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/31/2017
Initial Report FDA Received Date06/21/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age73 YR
-
-