Model Number H7493808008250
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/05/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that shaft break occurred.The 75% stenosed, 7x2.5mm target lesion was located in the moderately tortuous and moderately calcified left anterior descending artery (lad).A 2.5mm x 8mm balloon catheter was advanced for dilation.However, it was noted that the balloon shaft was fractured.The device was simply removed from the patient and the procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
 
Additional Manufacturer Narrative
Device evaluated by mfr: returned product consisted of a quantum maverick balloon catheter in two pieces.The shaft, hypotube, and bonds were microscopically and visually examined.There was contrast in the inflation lumen.The balloon was tightly folded.There were numerous kinks throughout the hypotube and shaft of the device.The hypotube had a complete hypotube separation 73cm from the strain relief.The hypotube fracture surfaces were ovaled, which suggests the device was kinked prior to separation.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
 
Event or Problem Description
It was reported that shaft break occurred.The 75% stenosed, 7x2.5mm target lesion was located in the moderately tortuous and moderately calcified left anterior descending artery (lad).A 2.5mm x 8mm balloon catheter was advanced for dilation.However, it was noted that the balloon shaft was fractured.The device was simply removed from the patient and the procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM¿ MAVERICK¿
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
MDR Report Key6657940
Report Number2134265-2017-06249
Device Sequence Number11145529
Product Code LOX
UDI-Device Identifier08714729392545
UDI-Public(01)08714729392545(17)20200131(10)20160491
Combination Product (Y/N)N
Initial Reporter CountryCN
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2020
Device Model NumberH7493808008250
Device Catalogue Number38080-0825
Device Lot Number20160491
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2017
Initial Date Received by Manufacturer 06/05/2017
Supplement Date Received by Manufacturer06/26/2017
Initial Report FDA Received Date06/21/2017
Supplement Report FDA Received Date07/10/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age69 YR
Patient Weight80
-
-