STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE PLUS TMZF HIP STEM #4; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 6021-0435
Device Problems Degraded (1153); Material Integrity Problem (2978); Naturally Worn (2988)
Patient Problems Pain (1994); Injury (2348)
Event Date 05/24/2017
Event Type  Injury  
Manufacturer Narrative
It was noted that the device is not available for evaluation.Additional information has been requested.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Event Description
Revision of a accolade i hip stem and 36mm +5 lfit head.Resmod stem, 32 ceramic head and x3 liner re-implanted.Patient came to doctor's office with complaint of hip pain.The trunnion was worn off like a pencil and notched.Therefore, the head wouldn't stay on and it wore a hole into the poly.This was the cause for the revision.This was a right hip revision.
 
Manufacturer Narrative
An event regarding wear involving an accolade stem was reported.The event was not confirmed.Method & results: device evaluation and results: a visual, functional and dimensional inspection could not be performed as the device was not returned.Medical records received and evaluation: no medical records or x-rays were made available for evaluation.Device history review: device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies complaint history review: there has been no other event for the lot referenced.Conclusions: the event could not be confirmed nor could the root cause be determined because the devices were not returned for evaluation and insufficient medical information was provided.If the devices and/or additional information are received, this investigation will be reopened and re-evaluated.
 
Event Description
Revision of a accolade i hip stem and 36mm +5 lfit head.Resmod stem, 32 ceramic head and x3 liner re-implanted.Patient came to doctor's office with complaint of hip pain.The trunnion was worn off like a pencil and notched.Therefore, the head wouldn't stay on and it wore a hole into the poly.This was the cause for the revision.This was a right hip revision.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE PLUS TMZF HIP STEM #4
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key6658568
MDR Text Key78125259
Report Number0002249697-2017-01958
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
PMA/PMN Number
K121308
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Type of Report Initial,Followup
Report Date 01/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/18/2013
Device Catalogue Number6021-0435
Device Lot Number26624606
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/24/2017
Initial Date FDA Received06/21/2017
Supplement Dates Manufacturer Received12/12/2017
Supplement Dates FDA Received01/10/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age76 YR
-
-