STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE PLUS TMZF HIP STEM #4; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
|
Back to Search Results |
|
Catalog Number 6021-0435 |
Device Problems
Degraded (1153); Material Integrity Problem (2978); Naturally Worn (2988)
|
Patient Problems
Pain (1994); Injury (2348)
|
Event Date 05/24/2017 |
Event Type
Injury
|
Manufacturer Narrative
|
It was noted that the device is not available for evaluation.Additional information has been requested.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
|
|
Event Description
|
Revision of a accolade i hip stem and 36mm +5 lfit head.Resmod stem, 32 ceramic head and x3 liner re-implanted.Patient came to doctor's office with complaint of hip pain.The trunnion was worn off like a pencil and notched.Therefore, the head wouldn't stay on and it wore a hole into the poly.This was the cause for the revision.This was a right hip revision.
|
|
Manufacturer Narrative
|
An event regarding wear involving an accolade stem was reported.The event was not confirmed.Method & results: device evaluation and results: a visual, functional and dimensional inspection could not be performed as the device was not returned.Medical records received and evaluation: no medical records or x-rays were made available for evaluation.Device history review: device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies complaint history review: there has been no other event for the lot referenced.Conclusions: the event could not be confirmed nor could the root cause be determined because the devices were not returned for evaluation and insufficient medical information was provided.If the devices and/or additional information are received, this investigation will be reopened and re-evaluated.
|
|
Event Description
|
Revision of a accolade i hip stem and 36mm +5 lfit head.Resmod stem, 32 ceramic head and x3 liner re-implanted.Patient came to doctor's office with complaint of hip pain.The trunnion was worn off like a pencil and notched.Therefore, the head wouldn't stay on and it wore a hole into the poly.This was the cause for the revision.This was a right hip revision.
|
|