Model Number ZCT300
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 05/24/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).All pertinent information available to the manufacturer has been submitted.
 
Event or Problem Description
It was reported that a zct300 22.0 diopter intraocular lens (iol) was explanted as the patient was experiencing glare around lights and was adjusting well to the lens.Surgery went successfully.No further information was available.
 
Additional Manufacturer Narrative
Device available for evaluation ¿ yes, returned to manufacturer on 07/13/2017.Device returned to manufacturer ¿ yes.Device evaluation: the device was returned to the manufacturer.Visual inspection with the unaided eye shows the product is cut in pieces, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.The customer's reported event could not be confirmed in the returned sample.Manufacturing records were reviewed and the lens was manufactured according to specification.A search on complaints revealed that no similar complaints were received for this production order number.The directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the device.Based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to the manufacturer has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
TORIC IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key6659033
Report Number9614546-2017-00551
Device Sequence Number13956614
Product Code HQL
UDI-Device Identifier05050474553361
UDI-Public(01)05050474553361(17)200413
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/16/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date04/13/2020
Device Model NumberZCT300
Device Catalogue NumberZCT300U220
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/13/2017
Initial Date Received by Manufacturer 05/30/2017
Supplement Date Received by Manufacturer07/24/2017
Initial Report FDA Received Date06/21/2017
Supplement Report FDA Received Date08/16/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-