• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ASCENSIA DIABETES CARE US INC. CONTOUR NEXT; BLOOD GLUCOSE TEST STRIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ASCENSIA DIABETES CARE US INC. CONTOUR NEXT; BLOOD GLUCOSE TEST STRIPS Back to Search Results
Model Number 7097C
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/07/2017
Type of Reportable Event Malfunction
Event or Problem Description
The customer received a blood glucose reading of 108mg/dl on the contour.He also tested on the contour usb and the readings were 67, 61 and 24mg/dl.The differences between the readings fall in the "c" and "d" zones of the consensus error grid, making the differences clinically significant.No adverse event was alleged.The customer was advised to return the test strips for investigation.New meter and strips were sent to him.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTOUR NEXT
Common Device Name
BLOOD GLUCOSE TEST STRIPS
Manufacturer (Section D)
ASCENSIA DIABETES CARE US INC.
430 s. beiger st.
mishawaka IN 46544
MDR Report Key6660897
Report Number1826988-2017-00209
Device Sequence Number16154185
Product Code NBW
UDI-Device Identifier40301937097506
UDI-Public40301937097506
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K062058
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 06/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/30/2018
Device Model Number7097C
Device Lot Number6LJ3A07B
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/07/2017
Initial Report FDA Received Date06/22/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-