| Model Number |
05-PA-65-40 |
| Medical Device Problem Code |
Disassembly (1168)
|
| Health Effect - Clinical Code |
Intraoperative Pain (2662)
|
| Date of Event |
05/25/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device was returned for evaluation and was found to be disassembled.When the components were inspected, they were found to be within specifications.Another screw of this lot was tested and the complaint failure mode could not be duplicated.The dhr was reviewed and the device was manufactured to pioneer surgical's specifications.
|
| |
|
Event or Problem Description
|
|
The patient had a posterior spinal rod fixation surgery on (b)(6) 2017.Approximately 2 weeks post op, the patient came in for a follow up visit and it was found that one of the pedicle screws had disassociated.The patient was revised and the screw was replaced.
|
| |
|