Model Number ETLW1628C124E
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Rupture (2208)
Date of Event 05/14/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Information references the main component of the system.Other relevant device(s) are: etlw1624c82e sn (b)(4), expiration date: 2016-05-15, udi # (b)(4).
 
Event or Problem Description
An endurant ii stent graft system was implanted in a patient for the endovascular treatment of an abdominal aortic aneurysm.It was reported that the patient presented emergently with a ruptured aneurysm with a type ii endoleak form a lumbar artery, type ia endoleak due to disease progression and type ib endoleak on the right limb.The decision was made to explant the stent graft and open repair.The physician stated that the cause of the event was because the patient was lost to follow-up and from the type ii endoleak.No additional clinical sequelae were reported and the patient is fine.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURANT II ILIAC STENT GRAFT
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key6661178
Report Number2953200-2017-01026
Device Sequence Number10809772
Product Code MIH
UDI-Device Identifier00613994991348
UDI-Public00613994991348
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P100021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/20/2016
Device Model NumberETLW1628C124E
Device Catalogue NumberETLW1628C124E
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/01/2017
Supplement Date Received by Manufacturer06/01/2017
Initial Report FDA Received Date06/22/2017
Supplement Report FDA Received Date10/04/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
-
-