A discordant, falsely low alpha-1-antitrypsin (aat) result was obtained on one patient sample on an advia 2400 instrument.The initial result was reported to the physician(s), who questioned it.The sample was repeated on an alternate advia chemistry instrument, resulting higher and matching the clinical picture of the patient.A new sample tube was obtained from the patient and was run on the alternate advia chemistry instrument, also resulting higher.The corrected result obtained from the new sample tube on the alternate advia chemistry instrument was reported to the physician(s).There are no reports of patient intervention or adverse health consequence due to the discordant, falsely low aat result.
|
The initial mdr 2432235-2017-00380 was filed on june 22, 2017.Additional information (07/07/2017): a siemens headquarters support center (hsc) specialist reviewed the information provided and concluded that there was no method or reagent issue.The hsc specialist stated that overflow of the dilution tray wash station would have diluted the sample and impacted result.The cause of the discordant, falsely low alpha-1-antitrypsin result on one patient sample is unknown.
|