Catalog Number TU-05500-J |
Device Problem
Infusion or Flow Problem (2964)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/07/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The device has not been returned for investigation at this time.Teleflex will continue to monitor and trend related events.
|
|
Event Description
|
When injecting the medical agent into the inserted catheter, the plunger of the syringe was so hard that the medical agent could not be injected through the catheter.After the catheter was replaced with a new one, the medical agent was re-injected using the same syringe but the same issue was duplicated.A new syringe was used and the therapy was continued without problem.There was no patient injury.
|
|
Manufacturer Narrative
|
(b)(4).The customer reported the medication would not inject through the catheter.The customer returned one snaplock adapter, one flat filter, and one epidural catheter.The components were received connected together.The returned components were visually examined with and without magnification.Visual examination of the returned snaplock adapter revealed that the snaplock appears typical with no observed defects or anomalies.Visual examination of the returned filter revealed that the filter appears typical with no observed defects or anomalies.Visual examination of the returned catheter revealed that the catheter appears typical but used.Biological material can be seen within the inner coils of the catheter especially at the distal end of the catheter.No other defects or anomalies were observed.A manual flow test was performed using the returned catheter, snaplock adapter, and filter.No blockages were found.Other remarks: a device history record review was performed on the epidural catheter with no relevant findings.A corrective action is not required at this time as there were no functional issues found with the returned sample.The reported complaint of medication not injecting into the catheter could not be confirmed based on the sample received.A device history record review was performed on the epidural catheter with no evidence to suggest a manufacturing related issue.The returned components passed a functional flow test.There were no functional issues found with the returned sample.Teleflex will continue to monitor and trend related events.
|
|
Event Description
|
When injecting the medical agent into the inserted catheter, the plunger of the syringe was so hard that the medical agent could not be injected through the catheter.After the catheter was replaced with a new one, the medical agent was re-injected using the same syringe but the same issue was duplicated.A new syringe was used and the therapy was continued without problem.There was no patient injury.
|
|
Search Alerts/Recalls
|