Catalog Number DR8054
Medical Device Problem Codes Break (1069); Leak/Splash (1354)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 05/16/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturing review: the device history records were reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Visual/microscopic inspection: the balloon size for this product is printed on the hub and identified the returned sample as a 5mm x 4cm balloon.Kinks were noted 32.0cm and 47.4cm from the distal tip.However after review of the preliminary pictures, the manner in which the device was packaged for return may have contributed to the kink.No kinks were reported by the user.No other anomalies were noted along the length of the catheter.Functional/performance evaluation: the patency of the guidewire lumen was tested using an in-house 0.035¿ guidewire, and the guidewire was able to pass the two kinks in the device catheter.The inflation hub was connected to an in-house inflation device and water was used to inflate the balloon.Upon inflation, water was observed leaking from the glue joint.The device was examined under magnification (40x) and a partial circumferential catheter break was identified at the glue joint.The balloon was unable to be inflated to rbp due to the leak.The balloon was able to be deflated without issue.Medical records review: medical records were not provided; therefore, a review could not be performed.Image/photo review: images/photos were not provided; therefore, a review could not be performed.Conclusion: an attempt was made to inflate the device, and a partial circumferential break was noted at the glue joint, confirming the investigation for the reported leak and for a break at the glue joint.The root cause for the partial break at the proximal balloon glue joint could not be determined based upon available information.It is unknown whether patient and/or procedural factors contributed to the event.Labeling review: the current ifu (instructions for use) states: warnings: if resistance is met during manipulation, determine the cause of the resistance before proceeding.Applying excessive force to the catheter can result in tip breakage or balloon separation.Use of the dorado pta dilatation catheter: advance the catheter through the introducer sheath and over the wire to the site of inflation.If the stenosis cannot be crossed with the desired dilatation catheter, use a smaller diameter catheter to pre-dilate the lesion to facilitate passage of a more appropriately sized dilatation catheter.Position the balloon relative to the lesion to be dilated, ensure the guidewire is in place and inflate the balloon to the appropriate pressure.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during preparation for an angioplasty procedure the catheter shaft allegedly leaked while attempting to flush the balloon.Another balloon was used to perform the procedure.There was no patient contact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORADO PTA BALLOON DILATATION CATHETER
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key6662280
Report Number2020394-2017-00697
Device Sequence Number13499687
Product Code DQY
UDI-Device Identifier00801741058493
UDI-Public(01)00801741058493(17)190802(10)93UA0165
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K072283
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/02/2019
Device Catalogue NumberDR8054
Device Lot Number93UA0165
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/24/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/26/2017
Initial Report FDA Received Date06/22/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-