ETHICON ENDO-SURGERY, LLC. ENDOPATH XCEL BLADELESS TROCAR; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY Back to Search Results
Catalog Number B12SRT
Device Problems Break (1069); Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/08/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The device history records were reviewed and the manufacturing criteria were met prior to the release of this lot.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.The following information was requested, but unavailable: is the stopcock missing? is the seal missing?.
 
Event Description
It was reported that during a laparoscopic myomectomy, it was found that the valve was missing when the forceps was inserted into the device.The broken pieces were retrieved and will be sent with the device.Another device was used to complete the case.There were no adverse consequences to the patient.
 
Manufacturer Narrative
(b)(4).Batch # n93f79, n93k3r, lot # p4r064.The device history records were reviewed and the manufacturing criteria were met prior to the release of this lot.The analysis results found that the b12srt device was returned with the duckbill damaged and torn; the torn piece of the duckbill was returned inside a petri dish.In addition, the tyvek was returned along with the instrument.The device was disassembled in order to evaluate the condition of the duckbill and the damage was confirmed; in addition, the separate piece of the duckbill was observed to have a mark which suggest that a pointy instrument was inserted through the trocar with excessive force.Per instructions for use: "use caution when introducing or removing instruments through the trocar sleeve in order to prevent inadvertent damage to the seals which could result in loss of pneumoperitoneum.Special care should be used when inserting sharp or angled edged endoscopic instruments to prevent tearing the seal." the batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
 
Manufacturer Narrative
(b)(4).Additional information was requested and the following was obtained: is the stopcock missing? no.Is the seal missing? yes.Did the seal fall into the patient? yes.How was the seal retrieved? no information how the seal was retrieved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOPATH XCEL BLADELESS TROCAR
Type of Device
LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo PR 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC
475 calle c
guaynabo PR 00969
Manufacturer Contact
milton garrett
4545 creek road ml 120a
cincinnati, OH 45242
5133378865
MDR Report Key6662907
MDR Text Key78375933
Report Number3005075853-2017-03223
Device Sequence Number1
Product Code GCJ
Combination Product (y/n)N
PMA/PMN Number
K032676
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 06/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/23/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2021
Device Catalogue NumberB12SRT
Device Lot NumberP4R064
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received06/28/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/09/2017
Is the Device Single Use? Yes
Type of Device Usage Initial