• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS TECNIS; TORIC IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS TECNIS; TORIC IOLS Back to Search Results
Model Number ZCT300
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/23/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If explanted, give date: not applicable as there is no indication that the lens was explanted.All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that a tecnis toric intraocular lens (iol) was implanted in the left eye of a (b)(6) male patient.The iol, initially placed at 72 degrees, was noticed rotated to 83 degrees, at one month post-op visit.Reportedly, the outcome does not significantly interferes with the patient's activities of daily life.No further actions were required.The patient was reported doing well.Bcdva (best corrected distance visual acuity) was 20/20 -2 os (ocular sinister - left eye).No further information was provided.
 
Additional Manufacturer Narrative
Device evaluation: the intraocular lens (iol) was not returned at the manufacturing site; therefore product testing could not be performed and the customer's reported complaint could not be verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search revealed that no similar complaints for this order number have been received.Labeling review: the directions for use (dfu) were reviewed.The directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to abbott medical optics has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
TORIC IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key6664437
Report Number9614546-2017-00562
Device Sequence Number19347837
Product Code HQL
UDI-Device Identifier05050474553309
UDI-Public(01)05050474553309(17)191215
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source health professional,study,use
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/03/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date12/15/2019
Device Model NumberZCT300
Device Catalogue NumberZCT300U190
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/01/2017
Supplement Date Received by Manufacturer07/12/2017
Initial Report FDA Received Date06/23/2017
Supplement Report FDA Received Date08/04/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age63 YR
-
-