• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APPLIED MEDICAL RESOURCES CTF71, KII FIOS 12X150MM Z-THREAD, 6BX; GCJ

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APPLIED MEDICAL RESOURCES CTF71, KII FIOS 12X150MM Z-THREAD, 6BX; GCJ Back to Search Results
Model Number CTF71
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/08/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The event device is anticipated to return.A follow-up report will be provided upon completion of the investigation.
 
Event or Problem Description
Procedure performed: robotic hysterectomy.Event description: "dr [doctor's name] was using the trocar as a robotic camera port.He removed the camera 1/4 of the way up the cannula and changed the angle of the camera.The distal end of the trocar must have been against another robotic arm/instrument and torqued the cannula causing it to crack.The broken piece did not detach and no patient injury occured." additional information received via email from territory manager on june 15, 2017 - davinci si robot was being used with no clamp.Type of intervention: na.Patient status: no patient injury or illness occurred that was associated with the complaint event.
 
Additional Manufacturer Narrative
The event product was returned to applied medical for evaluation.Upon inspection, engineering was able to confirm the complainant's experience of a fractured cannula.The wall thickness of the cannula was measured and was found to be uneven.The likely root cause of the fracture is the uneven cannula wall thickness, which occurred during the injection molding process of the cannula.The probability and criticality of harm resulting from this failure mode have been evaluated and were found to be at an acceptable level.Applied medical continuously seeks to improve the form, function and ease of use of its products.As part of this process, applied medical is currently researching possible process enhancements intended to further minimize the potential for this type of incident to occur.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CTF71, KII FIOS 12X150MM Z-THREAD, 6BX
Common Device Name
GCJ
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
MDR Report Key6664764
Report Number2027111-2017-01869
Device Sequence Number2059343
Product Code GCJ
UDI-Device Identifier00607915125035
UDI-Public(01)00607915125035(17)191216(30)01(10)1285762
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K041795
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date12/16/2019
Device Model NumberCTF71
Device Catalogue Number101470688
Device Lot Number1285762
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/29/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/09/2017
Supplement Date Received by Manufacturer06/09/2017
Initial Report FDA Received Date06/23/2017
Supplement Report FDA Received Date07/18/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-