MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD Back to Search Results
Catalog Number 400SMTXSFT4H12
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Infarction, Cerebral (1771); Vitreous Floaters (1866); Weakness (2145)
Date of Event 05/22/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The product was not returned for evaluation.From the information provided, there is no indication that there was any device malfunction, nonconformance, or misuse that contributed to the reported events.From the complaint description, the physician suspected that the eye floaters may have been transient due to the larger amount of contrast administered during the procedure than normal.Allergic reaction and anaphylaxis from contrast media, embolic stroke and other cerebral ischemic events are known and anticipated complications with these types of procedures and are noted in the device labeling.Therefore, the reported transient eye floaters and scattered punctuated infarct events are considered anticipated procedural complications.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report numbers: 3005168196-2017-01001, 3005168196-2017-01003, 3005168196-2017-01004, 3005168196-2017-01005, 3005168196-2017-01006, 3005168196-2017-01007, 3005168196-2017-01008.The device was implanted into the patient.
 
Event or Problem Description
On (b)(6) 2017, the patient underwent a coil embolization procedure in the anterior communicating artery using eight penumbra smart coils (smart coils) with no complications.Later that day, the patient was found to have transient eye floaters in the left eye of moderate severity.A computed tomography (ct) was taken and was negative for stroke.The physician suspected that the eye floaters may have been transient due to the larger amount of contrast administered during the procedure than normal.No other action was taken to treat this event.On (b)(6) 2017, the patient experienced neurological decline of right-sided weakness, resulting in prolonged hospitalization.A ct angiography was taken, which showed scattered punctuated infarcts throughout the bilateral frontal and parietal lobes suggestive of embolic disease.No action was taken to treat this event.The patient was discharged the next day on (b)(6) 2017.The reported transient eye floaters was adjudicated by the physician to be possibly related to the disease state, procedure, and the smart coil system.The reported neurological decline was adjudicated by the physician to be unrelated to the disease state, possibly related to the smart coil system, and definitely related to the procedure.As of (b)(6) 2017, both of the transient eye floaters and scattered punctuated infarct events are still ongoing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key6664865
Report Number3005168196-2017-01002
Device Sequence Number1970971
Product Code HCG
UDI-Device Identifier00814548017204
UDI-Public00814548017204
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/24/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Catalogue Number400SMTXSFT4H12
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/24/2017
Initial Report FDA Received Date06/23/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Life Threatening; Other;
Patient Age67 YR
-
-