| Model Number |
LDC315T |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/25/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The surgical center's material management reported an issue encountered while using the lacricath lacrimal duct catheter.The report stated that the device malfunctioned when the physician attempted to inflate the catheter's balloon.The physician stated that the device leaked fluid, so he removed the catheter and completed the procedure using a crawford tube instead.There were no patient complications reported as a result of the alleged issue.The device was returned to the manufacturer for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
The original report stated the device model to be ldc213t.Once the device was received and evaluated it was found to be model ldc315t.Device evaluation confirmed the reported complaint condition of the device leaking.Evaluation found that there was a tear in the balloon material.It is unknown where the tear originated from.There were no anomalies found in the dhr related to the reported complain condition.
|
| |
|