| Model Number |
GELWEAVE STRAIGHT |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
06/07/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient underwent an ecmo (extracorporeal membrane oxygenation) procedure at the time of the reported issue.No patient consequence was reported by the site.(b)(4).Device was not returned to vascutek ltd.For evaluation, and no devices from the applicable batch was available for evaluation.Results: the review of the retained quality, manufacturing and labelling records confirmed that the product was manufactured to the intended specifications.Unapproved or contraindicated for use - gelweave grafts are designed for "systemic vascular repair" procedures and the method of use as defined by the user (ecmo) is not indicated within the ifu; thus its use was off-label.No further actions are required however, the issue will be tracked and trended as part of the routine complaints monitoring and reporting process.If an adverse trend develops action may be taken at that time.Device not returned to the manufacturer.
|
| |
|
Event or Problem Description
|
|
Clinician had an issue with an 8mm gelweave graft on (b)(6) 2017 that was being used as part of an ecmo (extracorporeal membrane oxygenation) case.It was reported to be leaking throughout being described as a "soaker hose".The report also advised that the graft is not available for evaluation.
|
| |
|