Catalog Number 21-7231-24
Device Problem Partial Blockage (1065)
Patient Problem Hyperglycemia (1905)
Event Date 06/04/2017
Event Type  Injury  
Manufacturer Narrative
Customer has not returned the device to the manufacturer for device evaluation.If the device becomes available and is returned and evaluated, the manufacturer will file a follow-up report detailing the results of the evaluation.
 
Event Description
It was reported that during basal delivery using a cleo® 90 infusion set, the patient's pump displayed an occlusion alarm (alarm number in pump history: 26).Through troubleshooting with the distributor, it was determined that it was a likely there was a blockage located in the tubing.The patient's blood glucose levels were 444 mg/dl at the time of the event.To address the high blood glucose levels, the patient administered a manual insulin injection.The patient did not test for ketones.The patient was able to reattach the infusion set tubing to their site and successfully resumed insulin via the pump.No permanent injury was reported.See mfr: 3012307300-2017-01358.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEO® 90 INFUSION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
SMITHS HEALTHCARE MFG
s.a. de c.v. ave calidad no. 4
parque industrial internaciona
tijuana, 22425
MX   22425
Manufacturer Contact
jennifer meng
6000 nathan lane north
minneapolis, MN 55442
7633833078
MDR Report Key6666951
MDR Text Key78388619
Report Number3012307300-2017-01359
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10610586028403
UDI-Public10610586028403
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042172
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,distributor
Reporter Occupation Other
Type of Report Initial
Report Date 06/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/26/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/28/2021
Device Catalogue Number21-7231-24
Device Lot Number76X130
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received06/07/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/09/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NOVOLOG INSULIN
Patient Outcome(s) Required Intervention;
Patient Age22 YR
-
-