| Model Number |
ZKB00 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).All pertinent information available to the manufacturer has been submitted.
|
| |
|
Event or Problem Description
|
|
It was initially reported that a zkb00 5.0 diopter intraocular (iol) lens was going to be explanted due to a refractive error.Further information confirmed lens was explanted as planned on (b)(6) 2017.There was no incision enlargement, no vitrectomy and no sutures required.There was no patient post-op injury reported.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation ¿ yes, returned to manufacturer on 7/24/2017.Device returned to manufacturer ¿ yes.Device evaluation: the device was returned to the manufacturer.Visual inspection with the unaided eye shows the product is cut in pieces, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.The customer's reported event could not be confirmed in the returned sample.Manufacturing records were reviewed and the lens was manufactured according to specification.A search on complaints revealed that no similar complaints were received for this production order number.The directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the device.Based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to the manufacturer has been submitted.
|
| |
|
Search Alerts/Recalls
|