|
|
| Catalog Number |
AR-1927PSF |
| Health Effect - Clinical Codes |
Respiratory Distress (2045); Swelling (2091)
|
| Date of Event |
04/12/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Patient demographics (age at time of event, date of birth, gender, weight) were requested but not provided.No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.The device was not returned for evaluation but remained in the patient therefore the complainant's event could not be verified.The cause of the event could not be determined from the information available and without device evaluation.Device history record review revealed nothing relevant to this event.A possible cause of this type of event may be a reaction of the patient to the material(s) implanted or used during the implant procedure.Product directions for use warns of effects like foreign body and allergic-like reactions as well as infections both deep and superficial to the implant materials.Patient sensitivity to materials must be considered prior to implantation.This is the first complaint of this nature with this part/lot number combination.The potential cause(s) of this event will be communicated to the event reporter.If additional relevant information is received, a follow-up report will be submitted.This is one of five submissions for the same patient event.The others are 1220246-2017-00221 (cc116304, line 195529), 1220246-2017-00232 (cc116304, line 197799), 1220246-2017-00234 (cc116304, line 197801) and 1220246-2017-00235 (cc116304, line 197802).Device remains in patient.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent a rotator cuff repair on (b)(6) 2017 where a bio-composite swivelock tenodesis suture anchor, ar-1662bc-7 (cc116304, line 195529) was implanted into the patient's shoulder.Approximately four days after the surgery, the patient began to experience trouble breathing and swelling in the throat.The patient has been treated at the er for symptoms.No redness or swelling can be seen at the surgery site.The patient claims they have never had any issues of this nature until after their procedure.The patient's only known allergy is penicillin.The below part numbers were also implanted on (b)(6).Ar-1662bc lot 10110092 1ea (cc116304, line 197799), ar-1927psf lot 10068319 1ea (cc116304, line 197800), ar-1926ps lot 10064848 1ea (cc116304, line 197801) and ar-1926ps lot 10053400 1ea (cc116304, line 197802).
|
| |
|
Search Alerts/Recalls
|
|
|